Tempering barrel maintenance and maintenance (1) Add 300ml purified water to the inner barrel every week;
(2) It is forbidden to bring foreign matter or impurities into the inner barrel so as not to pollute the water quality and block the pipeline;
(3) After the experiment is completed, the constant temperature barrel lid should be opened to avoid corrosion of the instrument components;
(4) Replace the outer tub water at least once a year;
(5) Abnormal conditions of the instrument can be diagnosed using an expert diagnostic system.
Maintenance and Maintenance of Oxygen Bombs (1) Prior to the formal experiment every day, be sure to fill the empty oxygen bomb with enough gas and submerge it in water for several minutes to observe whether the sealing performance of the oxygen bomb is good and to ensure the reliability of the test results.
(2) After each experiment is completed, the parts of the oxygen bomb must be rinsed with distilled water and wiped with a special towel;
(3) If a leak of oxygen bomb is found, replace the spare seal; otherwise, it cannot be used;
(4) If it is found that the oxygen bomb thread slides, the use of the oxygen bomb is prohibited;
(5) Oxygen bombs should be inspected periodically and at least once a year for hydrostatic testing (which can be sent to Sande for testing);
(6) After use, it should be cleaned. It is best to heat on the electric oven for 3~5 minutes.
(7) Care must be taken when taking oxygen bombs from the inner barrel to prevent damage to the inner barrel due to impact or falling.
Automatic oxygenator maintenance and maintenance (1) The trial oxygen must meet the requirements of GB213-96, and the use of electrolytic oxygen is prohibited. Oxygen cylinders should be placed in a place that complies with safety regulations;
(2) The oxygenator should be placed on a stable work surface;
(3) Oxygenators (including oxygenating catheters and pressure reducing valves) must not be in contact with various oils and fats;
(4) The oxygen conduit on the oxygenator should avoid bending and twisting;
(5) There is no open flame around the oxygenator;
(6) If the oxygen pressure of the oxygen cylinder is less than 5 MPa, replace it with new oxygen. If the oxygen pressure in the oxygen bomb exceeds 3 MPa, the oxygen in the oxygen bomb should be released, and the oxygen bomb reloaded and recharged.
CNC busbar servo bending machine is the computer-controlled equipment dedicated for busbar bending, performing various bending operations through the coordination of X-axis and Y-axis movements, assisted with manual feeding. The selection of different dies can complete the plane bending, side bending and other processes of busbar. The machine can be used in combination with GJ3D software, and is able to accurately calculate the spread length of the bend. As to the frequently bent workpiece, it can automatically find the bending order to achieve programming automation, simulate the bending order to place busbar and avoid errors in determining the bending positions.